1. Product introduction
The YKSY800-2 Automatic 2 Positions Viscometer is suitable for polyester PET, PBT, nylon PA, PA6, PA66, etc.; polycarbonate PC, polyvinyl chloride PVC, polyacrylonitrile PAN, polyacrylamide PAM, polyvinylidene fluoride PVDF, organic silicon, organic glass PMMA, Polylactic acid PLA, PGA, cellulose CA, MCC, fiber paste and other polymer materials testing; executive standards: ISO1628, ASTMD4603, GB/T14190, ISO307, ASTMD789, GB/T12006, HB/T2234, GB/T3401, GB/ T12005, GB17514, GB/T28610, Pharmacopoeia and other international and domestic standards.
2. Product parameter
Model | YKSY800-2 |
Working temperature range | (normal temperature ~ 100)℃ |
Temperature control accuracy | ±0.01℃ |
Water temperature uniformity | ±0.01℃ |
Repeatability error | ±0.05s (Note: international, national standard±0.1s) |
Measuring unit stage difference | <0.1s |
Measurement time range | (0.001~999.999)s |
Timing accuracy | 0.001s or 0.1% of the measurement reading |
Viscosity range | 0.3-50000(mm2/s) |
Number of constant temperature tanks | 2 |
Flow through recirculating refrigerator | 400W |
Bit power | 1200W |
Power supply | 220V±5% 50Hz |
3. Product features
1. Measurement environment: The constant temperature tank adopts the design principle of stainless steel double chambers and a unique circulation method, and cooperates with the circulation of refrigeration through the refrigerator to ensure that the temperature of the entire measurement tank is uniform and the temperature stability of the fluctuation reaches±0.01°C. The drainage structure (water valve) is designed to facilitate cleaning and drainage.
2. Measurement module: Each measurement unit operates independently without interfering with each other, and the measurement unit can be operated at will, and the key parts are protected by corrosion-resistant materials. (Stand difference of each measuring unit: within 0.1s, keep the online cleaning module interface)
3. Measurement method: The measurement unit adopts two pairs of imported stainless steel optical fiber sensing devices, which ensure accurate detection. The sensing device can be automatically adjusted according to the color depth of the measured sample to meet the samples of different color depths, and has the characteristics of corrosion resistance, temperature resistance and long life.
4. Viscosity tube connection method: The cover method is adopted, which is easy to connect and operate, reduces the loss of the viscosity tube, and replaces the traditional (rubber cap and briquetting) viscosity tube connection method.
5. Viscosity measurement software: The software is compiled according to the industry, national and international standard methods, the user only needs to select the standard method to run, and the result will be calculated automatically. The software supports Win7 and above systems. The software is optimized and upgraded, and the measurement repeatability error is controlled within (±0.05s).
6. Safety: The YKSY800-2 Automatic 2 Positions Viscometer has over-temperature protection function and low water level alarm function, and the whole machine has leakage protection function.
4. Configuration list
No. | Item | Quantity |
1 | Viscosity precision constant temperature bath (2 positions) | 1 set |
2 | Circulating refrigerator (400W) | 1 set |
3 | Viscosity measuring | unit 2 |
4 | Main controller and software | 1 set |
5 | Baskets for viscosity tubes | 2 pcs |
6 | Ubbelohde viscosity tubes | 2 set |
7 | Viscometer bracket | 1 set |
8 | Large washers | 10 pcs |
9 | Small washers | 20 pieces |
10 | 0.01°C standard thermometer (23~26)°C | 1 piece |
11 | Lenovo Desktop PC | 1 set |
Remarks: The YKSY800-2 Automatic 2 Positions Viscometer can also be equipped with customized automatic sampling module (for sample injection), automatic cleaning module (for waste liquid discharge, cleaning and drying)