CS Analyzer
CS3500
CS3500 Series Analyzer is ideal to provide pricise and quick determination of carbon and sulfursimultaneously in primary steels, ores, finished metals, ceramics, and other inorganic materials.
CS3500 Series Analyzer is ideal to provide pricise and quick determination of carbon and sulfursimultaneously in primary steels, ores, finished metals, ceramics, and other inorganic materials.
CS3500 Series Analyzer is ideal to provide pricise and quick determination of carbon and sulfursimultaneously in primary steels, ores, finished metals, ceramics, and other inorganic materials.
A sample of typically 0.5 to 1.0g is weight into a crucible together with 1.5g of tungstenaccelerator. This crucible is placed on a pedestal which drives it into an induction furnace by apneumatic system. The sample is completely burned in the furnace by pure Oxygen.The Carbonof this sample is oxidized to CO2 and the Sulphur content to SO2. Dust and other particles aretrapped in a filter system.The CO2 and SO2 is transported from the furnace in the Oxygencarrier gas to the infrared adsorption detectors.The actual gas concentrations are determinedand the data are send to the Pc.The operation software is calculating the results afterintegration of the whole signal peaks. A calibration with known samples builds up a databasewhich created the necessary accuracy and precision. The burned crucible is driven out of thefurnace by the pneumatic system and disposed by an operator. The furnace and gas flowsystem is purged after analysis by clean Oxygen to avoid any interference (memory effect) ofthe individual samples.
The induction furnace ensures a fast and complete combustion at temperatures of about2200'c. lt’s robust desian ensures a long lifetime and minimized maintenance, With 18.5 MHzand 2.2 kVA its a suitable furnace for the combustion of a wide range of materials.
The unique design reduces a splattern and the dust from the combustion is trapped in an easyto clean metal filter.
The specialty for this application developed IR detector with up to 4 independent absorptionunits ensures the needed wide analytical range with high precision.The reliable design is knownfor long term stability and very low maintenance.
High tech detectors and interference filters are the base for a long time stability of thecalibration. A build in temperature controller makes the detector system independent from theenvironmental conditions.
Solid-state pyroelectric detector made in Germany
Flexible heating power control
*Advanced temperature control technology
*Advanced gas flow control technology
*Software has unique features
*IR signal data processing result is ace after linearizing
*Analytical precision reached international advance level
*Electric parts with high reliability
*Rich in diagnosis function
*Redundancy protection against meltdown
*Sealed gas passage with no dead volume
*Analysis range (500mg sample):
Low carbon:1ppm~0.1% high carbon:0.1%~5%
Low sulfur: 1ppm-0.3% high sulfur (150mg sample):0.3%-30%*
*Note:
The analysis range can be extended to 100% by changing the sample weight.
High sulfur absorption tank can be customized according to customer demand.
*Analysis accuracy:
Low carbon: 1ppm or RSD ≤ 1% high carbon: RSD ≤ 0.5%
Low sulfur: 1.5ppm or RSD s 1.5%, high sulfur: RSD s 1.5%
*Sensitivity:
0.01ppm
*Analysis time:
30s~40 s
Sample weighing:
The recommended range is 0.1g-0.5g, and the sample weight can be changed according to thesample content.
*High frequency induction furnace :
18mhz 2.7kva
*Power supply :
220 VAC + 10% 50 ±1 Hz, minimum current 16 A
*Instrument size :
55cm in width, 77cm in height and 76cm in thickness
*Weight:
100kg
*Carrier gas:
oxygen, purity 99.9%