1. Product introduction
This YK-3536-1 Cleveland Open Flash Point Tester is designed and manufactured in accordance with the requirements of the People's Republic of China Standard GB/T 3536-2008 "Determination of Flash Point and Ignition Point of Petroleum Products Cleveland Open Cup Method". Flash point and fire point of petroleum products and asphalt outside ℃.
2. Product parameter
Model | YK-3536-1 |
Working power supply | AC(220±10%)V, 50Hz |
Heating device | Electric furnace heating, no open flame, explosion-proof, power (0∼600) W continuously adjustable, maximum heating temperature 400°C |
Temperature control | Single-chip microcomputer temperature control, the heating rate meets the requirements of GB/T 3536-2008 standard |
Temperature display | The LCD screen displays various temperature parameters, the display value is (0-400) °C, and the accuracy is 0.1 °C |
Sweeping device | automatic sweeping |
Temperature sensor | PT100 platinum resistance temperature sensor |
Ignition device | (1) Ignition source: coal gas (or other civil combustible gas); (2) Nozzle aperture: about 0.8mm |
Ambient temperature | (-10∼50)℃ |
Relative humidity | ≤85% |
Power consumption of the whole machine | not more than 650W |
Dimensions | 340x320x450mm |
3. Product features
1. YK-3536-1 Cleveland Open Flash Point Tester adopts single-chip microcomputer control technology and LCD screen display, Chinese man-machine dialogue interface, continuous stepless adjustable heating power, and convenient parameter setting.
2. The liquid crystal display has a prompt menu and a prompt input operation interface, which can display the setting parameters and real-time display of the sample temperature and various test parameters. When the flash point appears, press the record key lightly, and the display will display and retain the flash point value.
3. The design is novel, the structure is compact, and the stainless steel table is beautiful. The user only needs to add gas or other civil combustible gas to carry out the test. It is easy to use and the test result is accurate.
4. YK-3536-1 Cleveland Open Flash Point Tester is equipped with a windshield and a flame extinguishing cover, which meets the requirements of the ignition test.
The biggest feature of this instrument is: the heating rate is accurate, only need to manually observe whether the flash point appears, the rest is done automatically by the instrument, the instrument is cost-effective.