Products Detail					Technical Parameter					Feature And Application																				Download				
				
															
						
1. Product introduction
					
					This YK-3536A Automatic Open Flash Tester is an automatic open flash point tester, which is an ideal substitute for similar imported instruments, and can be widely used in railway, aviation, electric power, petroleum industries and scientific research departments.
															
						
2. Product parameter
					
					| Model | YK-3536A | 
| Working power supply | AC220(-10%~+5%)V, 50Hz | 
| Temperature measurement | Range: room temperature to 400°C; Temperature display unit: 0.1°C | 
| Repeatability | ≤8°C (flash point and fire point) | 
| Reproducibility | ≤17°C (flash point), ≤14°C (ignition point) | 
| Heating speed | in line with GB/T3536 standard | 
| Ignition method | electronic ignition | 
| Ambient temperature | (10~40)℃ | 
| Relative humidity | ≤80% | 
| Power consumption of the whole machine | not more than 400W | 
| Dimensions | (when the lifting arm is not lifted) 410×360×310mm(When the lifting arm is raised) 410×360×420mm
 | 
															
						
3. Product features
					
					1. The YK-3536A Automatic Open Flash Tester adopts a large color LCD display, a touch-screen keyboard, and a full-Chinese man-machine dialogue interface. For parameters such as predictable flash point, sample label, atmospheric pressure, and test date, it has prompt menus and guided input functions. There are operating knowledge prompts on the screen.
2. Simulation tracking shows the function curve of temperature rise and test time. It has a small operation knowledge interface, Chinese misuse software prompt function, and can modify test date, test time and other parameters.
3. YK-3536A Automatic Open Flash Tester can store 10 screens and 120 sets of historical data.
4. No gas source is needed, and electronic ignition is adopted, which is safe to use and easy to operate.
5. Automatically correct the influence of atmospheric pressure on the test and calculate the correction value.
6. Differential detection, automatic correction of system deviation.
7. Swiping, testing, and printing data are automatically completed, and the test arm automatically rises and falls.
																														
						
Download
					
					
															keyword search:Automatic open flash tester